Abstract :
Comprehensive two-dimensional gas chromatography (GC × GC) has attracted much attention for the analysis of complex samples. Even with a large peak capacity in GC × GC, peak overlapping is often met. In this paper, a new method was developed to resolve overlapped peaks based on the mass conservation and the exponentially modified Gaussian (EMG) model. Linear relationships between the calculated σ, τ of primary peaks with the corresponding retention time (tR) were obtained, and the correlation coefficients were over 0.99. Based on such relationships, the elution profile of each compound in overlapped peaks could be simulated, even for the peak never separated on the second-dimension. The proposed method has proven to offer more accurate peak area than the general data processing method.
Keywords :
Comprehensive two-dimensional gas chromatography , EMG , Curve fitting , Overlapped peak , Deconvolution , Gas chromatography