• Title of article

    Calibration of a detector for nonlinear responses

  • Author/Authors

    Asnin، نويسنده , , Leonid and Guiochon، نويسنده , , Georges، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    6
  • From page
    105
  • To page
    110
  • Abstract
    A calibration curve is often needed to derive from the record of the detector signal the actual concentration profile of the eluate in many studies of the thermodynamic and kinetic of adsorption by chromatography. The calibration task is complicated in the frequent cases in which the detector response is nonlinear. The simplest approach consists in preparing a series of solutions of known concentrations, in flushing them successively through the detector cell, and in recording the height of the plateau response obtained. However, this method requires relatively large amounts of the pure solutes studied. These are not always available, may be most costly, and could be applied to better uses. An alternative procedure consists of deriving this calibration curve from a series of peaks recorded upon the injection of increasingly large pulses of the studied compound. We validated this new method in HPLC with a UV detector. Questions concerning the reproducibility and accuracy of the method are discussed.
  • Keywords
    Nonlinear calibration curve , HPLC , Reproducibility , UV-detector
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2005