A stability indicating assay method for cefuroxime axetil and its application to analysis of tablets exposed to accelerated stability test conditions
Author/Authors
Ivana، نويسنده , , Ivanovic and Ljiljana، نويسنده , , Zivanovic and Mira، نويسنده , , Zecevic، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
7
From page
209
To page
215
Abstract
Cefuroxime axetil is the esterified form of cefuroxime, injectable second generation cephalosporine antibiotic that can be given orally. Stereo and structural isomers of cefuroxime axetil (CA), anti-cefuroxime axetil (ACA) and Δ3-cefuroxime axetil (DCA), can be present in cefuroxime dosage forms as the process related impurities as well as possible degradation product. Sensitive and precise reversed-phase high-performance liquid chromatography (RP-HPLC) method was developed and validated for the determination of cefuroxime axetil in the presence of its degradation products in solid dosage forms. The RSD values for cefuroxime axetil, anti-cefuroxime axetil and Δ3-cefuroxime axetil of 1.80, 1.99 and 2.48%, respectively, indicated a good precision of the RP-HPLC method. Developed RP-HPLC method was sensitive with LOD = 0.08 μg mL−1 and LOQ = 0.60 μg mL−1 for anti-cefuroxime axetil and LOD = 0.06 μg mL−1 and LOQ = 0.45 μg mL−1 for Δ3-cefuroxime axetil. Holding studies were carried out on Ceroxim® tablets, according to ICH regulation at 30 °C/60% relative humidity (RH) and 40 °C/75% RH for 1, 2, 3 and 6 months. The review data from the stability studies conducted, show the significant content change of Δ3-cefuroxime axetil.
Keywords
Cefuroxime axetil , Degradation products , Geometrical and positional isomers , RP-HPLC , Stability testing
Journal title
Journal of Chromatography A
Serial Year
2006
Journal title
Journal of Chromatography A
Record number
1525478
Link To Document