Title of article
Determination of halogenated mono-alcohols and diols in water by gas chromatography with electron-capture detection
Author/Authors
Matthew ، نويسنده , , Brendan M. and Anastasio، نويسنده , , Cort، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
13
From page
65
To page
77
Abstract
We have developed an analytical method for the detection of halogenated alcohols in water with particular focus on 3-chloro-1,2-propanediol and 3-bromo-1,2-propanediol. In this method the target analytes are extracted from water, derivatized with heptafluorobutyric anhydride, and then analyzed with gas chromatography with electron-capture detection. The effects of water, pH and seawater constituents on the method were investigated. Method detection limits for a 5 ml aqueous sample ranged from 0.14 μg l−1 for 2-bromo-1,3-propanediol to 1.7 μg l−1 for 1,3-dichloro-2-propanol (1,3DCP).
Keywords
alcohols , Diols , Halogenated compounds , Halohydrins , Chloropropanediols , Bromopropanediols