Automated capillary gas chromatographic system to monitor ethylene emitted from biological materials
Author/Authors
Pham-Tuan، نويسنده , , Hai and Vercammen، نويسنده , , Joeri and Devos، نويسنده , , Christophe and Sandra، نويسنده , , Pat، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2000
Pages
11
From page
249
To page
259
Abstract
An automated capillary gas chromatographic system to measure ethylene emitted from biological materials is presented. The system consists of an on-line sampling device, a thermodesorption preconcentration apparatus and a capillary gas chromatograph with a flame ionization detection system. The limit of detection achievable on the GC system alone is 5 pg ethylene. The use of the strong Carboxen 1000 adsorbent at a sampling temperature as low as −50°C allows sampling of volumes up to a few liters. Ethylene concentrations at low ppt levels can be accurately and reproducibly determined.