• Title of article

    Technical note: Acquisition of CT models for radiotherapy applications with reduced tube heating

  • Author/Authors

    Balter، James M. نويسنده , , Lam، Kwok L. نويسنده ,

  • Issue Information
    دوفصلنامه با شماره پیاپی F1 سال 2015
  • Pages
    -58
  • From page
    59
  • To page
    0
  • Abstract
    The potential for changing computed tomography (CT) protocols to provide data sets that generate high quality digitally reconstructed radiographs (DRRs) from scans with very low tube currents is demonstrated. DRRs were generated from CT data acquired with slice thickness of 1, 3, and 5 mm, using high current to reduce noise in axial images. These DRRs were compared to one generated from a CT scan acquired using 1 mm aperture and very low (10 mA) current. The DRR generated via this technique is comparable to that generated with high current and 1 mm aperture, and higher resolution than from the 3 and 5 mm CT scans.