Title of article :
Beam orientation optimization in intensity-modulated radiation treatment planning
Author/Authors :
Pugachev، A. B. نويسنده , , Boyer، A. L. نويسنده , , Xing، L. نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
A method for designing a variable-SOBP (spread-out Bragg peak) ridge filter has been proposed. First, ridge filter parameters are determined by using a Monte Carlo calculation followed by a fast two-step iterative optimization. Then, tilting the ridge filter results in continuous variation of the SOBP width. Monte Carlo calculations show that depth dose uniformity changes from ±1.3% to ± 1.6% for SOBP widths ranging from 10.3 cm to 14.5 cm. Advantages of the proposed tilting ridge filter include a capability of continuous SOBP variation and cost-effective installation for a given SOBP width range.
Keywords :
inverse planning , treatment planning , backprojection , intensity modulation , beam orientation optimization
Journal title :
MEDICAL PHYSICS
Journal title :
MEDICAL PHYSICS