Title of article :
Halide ion-selective electrode array calibration
Author/Authors :
Baret، نويسنده , , M. and Massart، نويسنده , , D.L. and Fabry، نويسنده , , P. and Menardo، نويسنده , , C. and Conesa، نويسنده , , F.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 1999
Abstract :
The calibration of several ions (Cl−, Br−, F− and OH−) measured with an ion selective electrodes (ISE) array has been carried out in the presence of interferents using an experimental design and multivariate calibration methods. Partial least squares regression and principal component regression do not seem to improve the test set prediction compared to multivariate linear regression. In the case of very slight or no interference on the ISE, each ion can be determined using the corresponding ISE and univariate calibration methods, but the use of multivariate methods does not lead to worse results.
Keywords :
Multivariate calibration , Ion-selective electrodes , Univariate calibration , interferences