Title of article :
Flow injection determination of bromide ion in a developer using bromide ion-selective electrode detector
Author/Authors :
Masadome، نويسنده , , Takashi and Asano، نويسنده , , Yasukazu and Nakamura، نويسنده , , Takashi، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 1999
Pages :
6
From page :
595
To page :
600
Abstract :
A potentiometric flow injection determination method for bromide ion in a developer was proposed, by utilizing a flow-through type bromide ion-selective electrode detector. The sensing membrane of the electrode was Ag2S-AgBr membrane. The response of the electrode detector as a peak-shape signal was obtained for injected bromide ion in a developer. A linear relationship was found to exist between peak height and the concentration of the bromide ion in a developer in a concentration range from 1.0×10−3 to 1.0×10−2 mol l−1. The relative standard deviation for 10 injections of a 6×10−3 mol l−1 bromide ion in a developer was 1.3% and the sampling rate was ca 17–20 samples h−1. The present method was free from the interference of an organic reducing reagent, an organic substance in a developer sample solution for the determination of bromide ion in a developer.
Keywords :
Bromide-selective electrode detector , Flow injection analysis , DEVELOPER , Bromide ion
Journal title :
Talanta
Serial Year :
1999
Journal title :
Talanta
Record number :
1639473
Link To Document :
بازگشت