Title of article
Application of neural network calibrations to an halide ISE array
Author/Authors
Baret، نويسنده , , M and Massart، نويسنده , , D.L. and Fabry، نويسنده , , P and Conesa، نويسنده , , F and Eichner، نويسنده , , C and Menardo، نويسنده , , C، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2000
Pages
15
From page
863
To page
877
Abstract
The Neural Network (NN) technique was applied to the calibration of an ion selective electrode (ISE) array comprising a bromide selective electrode, two chloride ISEs and one thiocyanate ISE. The measured samples were synthetic mixture solutions of chlorides and bromides in concentration ranges such that interference occurs. The NN method allowed to perform the calibration without estimating the coefficients of the Nikolskii–Eisenman theoretical relation. Only the determination of bromide was detailed. The results obtained using this method were better than those obtained using linear multivariate calibration methods.
Keywords
neural network , Bromide , ISEs , Interference
Journal title
Talanta
Serial Year
2000
Journal title
Talanta
Record number
1640086
Link To Document