Title of article :
Potentiometric flow injection analysis of concentrated hydrogen peroxide by using an Fe(II)–Fe(III) redox potential buffer solution
Author/Authors :
Imato، نويسنده , , Toshihiko and Ohura، نويسنده , , Hiroki and Yamasaki، نويسنده , , Sumio and Asano، نويسنده , , Yasukazu، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2000
Pages :
8
From page :
19
To page :
26
Abstract :
The flow injection analysis of hydrogen peroxide is proposed, using a redox electrode and an Fe(II)–Fe(III) potential buffer solution. Influencing factors, such as the concentrations of Fe(II)–Fe(III) and sulfuric acid in the potential buffer on sensitivity of the proposed method are examined. The analysis of high concentrations of hydrogen peroxide up to ∼10 M was conducted successfully with relative standard deviation of 0.7%.
Keywords :
Hydrogen peroxide , Potentiometric determination , Flow injection analysis , Iron(II)–iron(III) potential buffer
Journal title :
Talanta
Serial Year :
2000
Journal title :
Talanta
Record number :
1640225
Link To Document :
بازگشت