Title of article
Application of an ion-selective field effect transistor with a photocured polymer membrane in nephrology for determination of potassium ions in dialysis solutions and in blood plasma
Author/Authors
Abramova، نويسنده , , N and Borisov، نويسنده , , Yu and Bratov، نويسنده , , A and Gavrilenko، نويسنده , , P and Dom??nguez، نويسنده , , C and Spiridonov، نويسنده , , Elena Suglobova، نويسنده , , E، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2000
Pages
6
From page
533
To page
538
Abstract
Application of a potassium ion sensor based on an ion sensitive field effect transistor (ISFET) for ion control of a dialysis solution in an artificial kidney and in blood plasma of patients treated by hemodialysis is presented. Sensors and their long-term stability were characterised in constant contact with test solutions. Test results are compared to those obtained with conventional ion-selective electrodes and commercial blood ion analyser. Tested ISFET sensors showed high reliability in potassium ion measurements in the physiologically significant concentration range which, along with low cost of their production, makes them promising for cited application.
Keywords
Potassium ion sensor , Photocured polymer , ISFET , Artificial kidney
Journal title
Talanta
Serial Year
2000
Journal title
Talanta
Record number
1640434
Link To Document