Title of article :
Flow injection determination of iodide ion in a photographic developing solution using iodide ion-selective electrode detector
Author/Authors :
Masadome، نويسنده , , Takashi and Sonoda، نويسنده , , Risa and Asano، نويسنده , , Yasukazu، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2000
Abstract :
A potentiometric flow injection determination method for iodide ion in a photographic developing solution was proposed by utilizing a flow-through type iodide ion-selective electrode detector. The sensing membrane of the electrode was Ag2S–AgI membrane. The response of the electrode detector as a peak-shape signal was obtained for injected iodide ion in a photographic developing solution. A linear relationship in the subnernstian zone was found to exist between peak height and the concentration of the iodide ion in a photographic developing solution in a concentration range from 0 to 6.0×10−5 mol l−1. The relative standard deviation for ten injections of 2×10−5 mol l−1 iodide ion in a photographic developing solution was 0.96% and the sampling rate was approximately 12–13 samples h−1. The iodide ion could be determined under coexisting of an organic reducing reagent and inorganic electrolytes of high concentration in a photographic developing solution sample solution by the present method.
Keywords :
Photographic developing solution , Iodide-selective electrode detector , Flow injection analysis , Iodide ion