Title of article
Amperometric biosensor for the quantification of gentisic acid using polyphenol oxidase modified carbon paste electrode
Author/Authors
Pedano، نويسنده , , M.L and Rivas، نويسنده , , G.A، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2000
Pages
7
From page
489
To page
495
Abstract
The affinity of mushroom polyphenol oxidase (PPO) towards gentisic acid (GA), a metabolite of acetyl salicylic acid (ASA), is demonstrated by spectrophotometry and by electrochemical techniques. The enzyme can selectively recognize GA even in the presence of large excess of ASA and its metabolic derivatives (salicylic acid (SA) and salicyluric acid (SUA)). At −0.150 V, the sensitivity is (6.1±0.1)×104 NAM−1, the response is linear up to 2.0×10−4 M and the detection limit is 5.0×10−5 M. The kinetic parameters, obtained from Eadie–Hofstee plots, are Imax=51.4 nA and Kmapp=6.7×10−4 M.
Keywords
Enzymatic electrode , Amperometric biosensor , gentisic acid , Polyphenol oxidase , Carbon paste