Title of article :
Chemiluminescence sensor for isoniazid with controlled-reagent-release technology
Author/Authors :
Song، نويسنده , , Zhenghua and Lü، نويسنده , , Jihong Solomon Zhao، نويسنده , , Tiezhu، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2001
Pages :
7
From page :
1171
To page :
1177
Abstract :
A novel chemiluminescence (CL) sensor for isoniazid combined with flow-injection technology is presented in this paper. The analytical reagents, luminol and ferricyanide, were both immobilized on an anion-exchange column. The CL signal produced by the reaction between luminol and ferricyanide, which were eluted from the column through sodium phosphate injection, was decreased in the presence of isoniazid. The decreased CL intensity was linear with isoniazid concentration in the range 0.001–1.0 μg·ml−1; and the detection limit was 0.35 ng·ml−1 (3s). The whole process, including sampling and washing, could be completed in 2 min with a relative standard deviation of less than 4.1%. The sensor could be reused more than 400 times and has been applied for the determination of isoniazid in pharmaceutical preparations.
Keywords :
Luminol , Isoniazid , Sensor , Chemiluminescence , Immobilized reagents