• Title of article

    Optimization of experimental parameters for the determination of mercury by MIP/AES

  • Author/Authors

    Murillo، نويسنده , , Miguel and Carriَn، نويسنده , , Nereida and Chirinos، نويسنده , , José and Gammiero، نويسنده , , Alexis and Fassano، نويسنده , , Edmondo، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2001
  • Pages
    7
  • From page
    389
  • To page
    395
  • Abstract
    A system using a microwave induced plasma, as an emission source was optimized for total mercury determinations. The system contains a flow injection section in which mercury is reduced and deposited on a gold/platinum collector, and carried to the emission source. The microwave was generated in a Surfatron cavity. An experimental design was applied to optimize parameters concerning plasma conditions, amalgamation, desorption and mercury vapor generation. The limit of detection of the method was 14 pg ml−1 of mercury, using peak area mode while LDs was 1.3 pg ml−1 using peak height mode, with a 2% RSD and a linear dynamic range of almost three orders of magnitude. The accuracy was assessed using a certified reference material of atmospheric particulate National Institute of Standards and Technology (NIST) 2704. No significant difference at 95% confidence level was observed between the certified value and our result.
  • Keywords
    Mercury determination , Microwave Plasma MIP , Surfatron
  • Journal title
    Talanta
  • Serial Year
    2001
  • Journal title
    Talanta
  • Record number