• Title of article

    Flow-injection system for automated dissolution testing of isoniazid tablets with chemiluminescence detection

  • Author/Authors

    Li، نويسنده , , Baoxin and Zhang، نويسنده , , Zhujun and Liu، نويسنده , , Wei، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2001
  • Pages
    6
  • From page
    697
  • To page
    702
  • Abstract
    A simple and sensitive flow-injection chemiluminescence (CL) system for automated dissolution testing is described and evaluated for monitoring of dissolution profiles of isoniazid tablets. The undissolved suspended particles in the dissolved solution were eliminated via on-line filter. The novel CL system of KIO4-isoniazid was also investigated. The sampling frequency of the system was 120 h−1. The dissolution profiles of isoniazid fast-release tablets from three sources were determined, which demonstrates the stability, great sensitivity, large dynamic measuring range and robustness of the system.
  • Keywords
    Chemiluminescence , Flow-injection analysis , Isoniazid , dissolution testing
  • Journal title
    Talanta
  • Serial Year
    2001
  • Journal title
    Talanta
  • Record number