Title of article :
Detection of phenolic compounds in flow systems based on tyrosinase-modified reticulated vitreous carbon electrodes
Author/Authors :
Peٌa، نويسنده , , N. and Reviejo، نويسنده , , A.J. and Pingarrَn، نويسنده , , J.M.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2001
Abstract :
The fabrication and performance of a reticulated vitreous carbon (RVC)-based tyrosinase flow-through electrode, in which the enzyme was covalently immobilized, is reported. The bioelectrode was tested as an amperometric detector for phenolic compounds. Variables affecting the construction of the enzyme flow-through electrode such as the RVC chemical pretreatment procedure, the enzyme immobilization method in the RVC matrix, the enzyme loading and the pH value of the buffer solution used, were optimized by flow-injection with amperometric detection. A good immobilization of the enzyme in the RVC matrix, in spite of the hydrodynamic conditions, was found. The same tyrosinase-RVC electrode could be used with no significant loss of the amperometric response for around 20 days, and reproducible responses could be achieved with different electrodes constructed in the same manner. Moreover, the operational stability of the bioelectrode was tested under continuous monitorization conditions. Calibration plots by flow injection with amperometric detection at −0.20 V were obtained for phenol, 2,4-dimethylphenol; 3-chlorophenol; 4-chlorophenol; 4-chloro-3-methylphenol and 2-aminophenol, with detection limits ranging from 2 μg l−1 (4-chloro-3-methylphenol) to 2 mg l−1.
Keywords :
tyrosinase , Phenolic compounds. , flow-through , Reticulated vitreous carbon