Title of article :
Determination of trace hydrazine by differential pulse voltammetry using magnetic microspheres
Author/Authors :
Yang، نويسنده , , Ming and Li، نويسنده , , Hu Lin، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2001
Pages :
6
From page :
479
To page :
484
Abstract :
A new type of magnetic polymer microspheres containing ketone groups on the surface was synthesized. It can be reacted with hydrazine to produce electroactive adduct. Reduction of this derivative that was aggregated on the magnetic electrode is possible and effective in indirect determination of hydrazine. The experimental conditions and electrode structure were discussed. Under optimum conditions, It was found that the peak potential (Ep) of hydrazine is −1.06 V (vs. Ag/AgCl). Hydrazine in the range 0.3–500 μg l−1 can be determined. The detection limits for hydrazine is 0.1 μg l−1. The relative standard deviation for determination of 100 μg l−1 hydrazine was 2.43 %. The method was applied to the determination of hydrazine in water samples with satisfactory results.
Keywords :
Magnetic microsphere , hydrazine , Differential pulse voltammetry , Magnetic electrode
Journal title :
Talanta
Serial Year :
2001
Journal title :
Talanta
Record number :
1641777
Link To Document :
بازگشت