Title of article :
Copper(II)-selective membrane electrodes based on some recently synthesized mixed aza-thioether crowns containing a 1,10-phenanthroline sub-unit
Author/Authors :
Shamsipur، نويسنده , , Mojtaba and Javanbakht، نويسنده , , Mehran and Mousavi، نويسنده , , Mir Fazlollah and Ganjali، نويسنده , , Mohammad Reza and Lippolis، نويسنده , , Vito and Garau، نويسنده , , Alessandra and Tei، نويسنده , , Lorenzo، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2001
Abstract :
Three different mixed aza-thioether crowns containing a 1,10-phenanthroline sub-unit were investigated to characterize their abilities as copper(II) ion carriers in PVC-membrane electrodes. The electrode based on L1 exhibited a Nernstian response for Cu2+ ions over a wide concentration range (2×10−1 to 1×10−5 M) with a limit of detection of 8.0×10−6 M (0.5 p.p.m.). The response time of sensor is 15 s, and the membrane can be used for more than 3 months without observing any deviation. The electrode revealed comparatively good selectivities with respect to many alkali, alkaline earth, transition and heavy metal ions, and could be used in a pH range of 2.5–5.5. It was applied to the direct determination and potentiometric titration of the copper(II) ion.
Keywords :
PVC Membrane , Potentiometry , Copper ion-selective electrode , Mixed aza-thiother crowns