Title of article :
Flow-injection method for the determination of serine using immobilized enzyme
Author/Authors :
Yaqoob، نويسنده , , M. and Nabi، نويسنده , , A.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2001
Pages :
6
From page :
1181
To page :
1186
Abstract :
A flow-injection method for the determination of serine using a mini-column containing immobilized serine dehydratase isolated and purified from rat liver is described. Ammonia produced from the enzymatic reaction is coupled with hypochlorite and phenol in an alkaline medium yields a blue product due to the indophenol anion formation, which is the basis of a spectrophotometric detection at 640 nm. The limit of detection (2×blank noise) is 0.01 mM with a sample throughput of 25 h−1. Calibration graph is linear in the range 0.2–1.0 mM, with relative standard deviation 0.6–1.0%.
Keywords :
serine , Serine dehydratase , Flow-injection analysis , Spectrophotometer , Immobilized enzyme