Title of article :
Carbon fiber cylindrical microelectrode-based detector for the determination of antithyroid drugs
Author/Authors :
Guzm?n، نويسنده , , Alicia and Agü??، نويسنده , , Lourdes and Pedrero، نويسنده , , Mar??a and Y??ez-Sede?o، نويسنده , , Paloma and Pingarr?n، نويسنده , , José M، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2002
Pages :
8
From page :
577
To page :
584
Abstract :
An amperometric flow-injection method for the determination of antithyroid drugs such as 6-methyl-2-thiouracil (MTU) using a carbon fiber cylindrical microelectrode-based detector is reported. A home-made flow-cell specially adapted for working with cylindrical microelectrodes ranging between 4 and 10 mm in length was used for this purpose. Methanol containing 0.05 mol l−1 tetrabutylammonium perchlorate as the supporting electrolyte was employed, and a potential of +1.6 V was selected for the amperometric detection of MTU. No cleaning or electrode surface regeneration of the fiber was necessary during the whole working day. A limit of detection of 2.6×10−7 mol l−1 (37 μg MTU l−1) was achieved. The method was applied to the determination of MTU in spiked feed samples (peas and corn flour), at the 142 μg g−1 level. The procedure involved the extraction of the drug in methanol, a clean-up step using Florisil cartridges and the use of the standard additions method. Recoveries of 81±6 and 85±8% were obtained for peas and corn flour, respectively.
Keywords :
Flow-injection , Antithyroid drugs , Microelectrodes
Journal title :
Talanta
Serial Year :
2002
Journal title :
Talanta
Record number :
1642204
Link To Document :
بازگشت