Title of article
About pycnometric density measurements
Author/Authors
Viana، نويسنده , , C. Jouannin، نويسنده , , P and Pontier، نويسنده , , C and Chulia، نويسنده , , D، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2002
Pages
11
From page
583
To page
593
Abstract
Pycnometric density is at the moment the closest approximation of true density calculated from the molecular weight and crystalline lattice of the product. It is determined by using helium pycnometers that offer the advantage of being easy-to-use and rapid, especially fully automated apparatus. If the accuracy and the reproducibility of the technique are sufficient to reveal minute variations, this data is of interest for characterisation of crystalline structures (polymorphs, pseudopolymorphs, amorphous state), detection of defects, pores or impurities, and possible changes in crystal density during compaction. The aim of this article is to investigate on the confidence that can be expected when measuring density with an AccuPyc 1330 pycnometer. The experiments carried out on glass beads, quartz and mannitol have shown that operating parameters can affect the accuracy of pycnometric density to the nearest 0.01 g cm−3. If measurements are performed in optimal conditions, 0.02% accuracy can be obtained, otherwise it can fail to 0.1% or less, depending on the material and the variations in the standard volume.
Keywords
Validation , Pycnometric density , characterisation
Journal title
Talanta
Serial Year
2002
Journal title
Talanta
Record number
1642604
Link To Document