Title of article :
Determination of arsenic at a gold-plated carbon paste electrode using constant current stripping analysis
Author/Authors :
?vancara، نويسنده , , Ivan and Vyt?as، نويسنده , , Karel and Bobrowski، نويسنده , , Andrzej and Kalcher، نويسنده , , Kurt، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2002
Abstract :
A new stripping method for the determination of arsenic in water samples with a gold film-plated carbon paste electrode has been developed for the use in constant current stripping analysis (CCSA). In the novelized procedure, differentiation between As(III) and chemically pre-reduced As(V), the effect of Cu(II) on the response of arsenic, and the stability of sample solutions were studied in detail. Compared to the voltammetric approach, the method utilizing CCSA offers a more rapid procedure with improved analytical characteristics such as reproducibility, selectivity over the Cu(II) ions, or lower detection limit (3 ppb for As(III) and 0.5 ppb for As(V), respectively). The possibilities of the optimized method are demonstrated by determinations of As(III), As(V), and total arsenic in samples of polluted river water.
Keywords :
Constant current stripping analysis , Arsenic(III) and arsenic(V) determination and differentiation , River water , Gold-plated carbon paste electrode