• Title of article

    Construction and evaluation of a flow-through cell adapted to a commercial static mercury drop electrode (SMDE) to study the adsorption of Cd(II) and Pb(II) on vermiculite

  • Author/Authors

    Abate، نويسنده , , Gilberto and Lichtig، نويسنده , , Jaim and Masini، نويسنده , , Jorge C، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2002
  • Pages
    11
  • From page
    433
  • To page
    443
  • Abstract
    This paper describes the construction and application of a robust flow-through cell for use with the capillary of a commercial static mercury drop electrode. Linearity of peak current was observed up to 0.50 μmol l−1 for Cd(II) or Pb(II) in anodic stripping voltammetry experiments performed under continuous flow during the deposition step, using 120 s of deposition time and flow rate of 4.0 ml min−1. Under these conditions the limits of detection for Cd(II) and Pb(II) were 13 and 17 nmol l−1, respectively. An analytical throughput of 20 analyses per h was possible using 10 s for cleaning the cell between two samples and including the time needed for the potential scan, which was performed with the flow stopped, using the differential pulse mode for current sampling. The linear dynamic range can be extended up to 5 μmol l−1 for both cations if the deposition time is decreased to 30 s, a condition in which the sampling throughput is 35 analyses per h. The proposed manifold was used to study the adsorption rates of Cd(II) and Pb(II) onto vermiculite at different pHs, allowing one to perform high sensitivity measurements at high sampling frequency, using low cost instrumentation.
  • Keywords
    Continuous flow analysis , Adsorption , clays , Heavy metals , Voltammetry
  • Journal title
    Talanta
  • Serial Year
    2002
  • Journal title
    Talanta
  • Record number

    1643259