Title of article
Cesium ion selective electrode based on calix[4]crown ether–ester
Author/Authors
Mahajan، نويسنده , , R.K. and Kumar، نويسنده , , Manov and Sharma (nee Bhalla)، نويسنده , , V and Kaur، نويسنده , , Inderpreet، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2002
Pages
6
From page
445
To page
450
Abstract
The potentiometric response characteristics of cesium ion selective PVC membrane electrode employing calix[4]crown ether–ester as an ionophore were investigated. The electrode exhibit a good response for cesium ion over wide concentration range of 5.0×10−6–1.0×10−1 M with a Nernstian slope of 59 mV per decade. The detection limit of electrode is 5.0×10−6 M. The electrode was found to have selectivity for cesium ion over alkali, alkaline and transition metals. The response time of the electrode is less than 20 s and can be used for more than 4 months without observing any divergence in potentiometric response. The electrode response was stable over wide pH range.
Keywords
Cesium ion , Sensor
Journal title
Talanta
Serial Year
2002
Journal title
Talanta
Record number
1643262
Link To Document