Title of article
A flow cell optosensor for lead based on immobilized gallocynin in chitosan membrane
Author/Authors
Yusof، نويسنده , , Nor Azah and Ahmad، نويسنده , , Musa، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2002
Pages
8
From page
459
To page
466
Abstract
Gallocynin immobilized in chitosan membrane has been studied as a sensor element of an optical sensor for lead using a flowing system. By using this set up, lead in solution has been determined in the concentration range from 1.0×10−1 to 1.0×103 ppm with a detection limit of 0.075 ppm. The standard deviation of the method for the repeatability of lead detection at a concentration of 100 ppm was found to be 2.10%. The response of the sensor was reproducible and can be regenerated by using acidified saturated KNO3 solution. Interference from foreign ions was also studied at 1:1 mole ratio of Pb(II):foreign ions.
Keywords
Flow injection analysis (FIA) , Pb(II) sensor , Chitosan membrane , Optical fibre
Journal title
Talanta
Serial Year
2002
Journal title
Talanta
Record number
1643268
Link To Document