Title of article :
Flow-injection preconcentration and graphite furnace atomic absorption spectrometric determination of platinum
Author/Authors :
Lلsztity، نويسنده , , A and Kelkَ-Lévai، نويسنده , , ء and Zih-Perényi، نويسنده , , K and Varga، نويسنده , , I، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2003
Abstract :
Flow-injection graphite furnace atomic absorption spectrometric (GFAAS) methods were worked out using oxime, sulphoxine and 2,2′-diamino-diethylamine (DEN) cellulose microcolumns for preconcentration of platinum after reduction by iodide or sulphite ions. The detection limits were, at 20-fold enrichment, 0.21, 0.18 and 0.30 μg l−1, respectively. The total reflection X-ray fluorescence spectrometry (TXRF) was also used for the determination of platinum in eluates. The method was applied for the determination of platinum in salmeterol xinafoate and Ca-folinate pharmaceutical compounds. Decomposition of organic matrix of Ca-folinate was necessary before the preconcentration.
Keywords :
IV) preconcentration , Platinum(II , Pharmaceuticals , Chelating cellulose exchangers , Atomic spectrometry