Title of article :
Chemiluminescence microfluidic system sensor on a chip for determination of glucose in human serum with immobilized reagents
Author/Authors :
Lv، نويسنده , , Yi and Zhang، نويسنده , , Zhujun and Chen، نويسنده , , Funan، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2003
Abstract :
A chemiluminescence (CL) biosensor on a chip coupled to microfluidic system is described in this paper. The CL biosensor measured 25×45×5 mm in dimension, was readily produced in analytical laboratory. Glucose oxidase (GOD) was immobilized onto controlled-pore glass (CPG) via glutaraldehyde activation and packed into a reservoir. The analytical reagents, including luminol and ferricyanide, were electrostatically co-immobilized on an anion-exchange resin. The most characteristic of the biosensor was to introduce the air as the carrier flow in stead of the common solution carrier for the first. The glucose was sensed by the CL reaction between hydrogen peroxide produced from the enzymatic reaction and CL reagents, which were released from the anion-exchange resin. The proposed method has been successfully applied to the determination of glucose in human serum. The linear range of the glucose concentration was 1.1–110 mM and the detection limit was 0.1 mM (3σ).
Keywords :
Microfluidic system , Sensor , Chemiluminescence , Glucose , CHIP