    A simple, particle size-independent spectrometric method has been developed for the multi-element determination of wear metals in used engine oil. A small aliquot (0.5 ml) of an acid-digested oil sample is spotted onto a C-18 solid phase extraction disk to form a uniform thin film. The dried disk is then analyzed directly by energy dispersive X-ray fluorescence spectrometry. This technique provides a homogeneous and reproducible sample surface to the instrument, thus overcoming the typical problems associated with uneven particle size distribution and sedimentation. As a result, the method provides higher precision and accuracy than conventional methods. Furthermore, the disk sample may be stored and re-analyzed or extracted at a later date. The signals arising from the spotted disks, and the calibration curves constructed from them, are stable for at least 2 months. The limits of detection for Fe, Cu, Zn, Pb, and Cr are 5, 1, 4, 2, and 4 μg g−1, respectively. Recoveries of these elements from spiked oil samples range from 92 to 110%. The analysis of two standard reference materials and a used oil sample produced results comparable to those found by inductively coupled plasma atomic emission spectrometry.
  • Keywords
    X-ray fluorescence spectrometry , Energy dispersive , Inductively coupled plasma atomic emission spectrometry , Microwave digestion , Wear metals , Used engine oil
  • Journal title
    Talanta
  • Serial Year
    2003
  • Journal title
    Talanta
  • Record number

    1643820