• Title of article

    Determination of cyanide using flow-injection multisensor system

  • Author/Authors

    Ipatov، نويسنده , , Andrey and Ivanov، نويسنده , , Maxim and Makarychev-Mikhailov، نويسنده , , Sergey and Kolodnikov، نويسنده , , Vasily and Legin، نويسنده , , Andrey and Vlasov، نويسنده , , Yuri، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2002
  • Pages
    6
  • From page
    1071
  • To page
    1076
  • Abstract
    A flow-injection multisensor system (FIMS) comprising potentiometric sensors of different types for determination of free cyanide activity in basic solutions for extraction of noble metals was developed. The solvent polymeric membrane sensors based on metalloporphyrin and crystalline sensors were combined in the sensor system. The system allowed determination of cyanide activity in the range 10−4–1 mol l−1 with an error less than 5% in individual cyanide solutions and acceptable precision (about 20%) in process liquids. The system was able to analyse up to 20 samples per h. The FIMS was also applied to detecting of silver ions in the presence of cyanide. Chalcogenide glass sensor was used as the detector that ensured the precision of 20%.
  • Keywords
    Multisensor system , Cyanide determination , Flow-injection analysis
  • Journal title
    Talanta
  • Serial Year
    2002
  • Journal title
    Talanta
  • Record number