Author/Authors :
Ermolenko، نويسنده , , Yu. and Yoshinobu، نويسنده , , T. and Mourzina، نويسنده , , Yu. and Furuichi، نويسنده , , K. and Levichev، نويسنده , , S. and Schِning، نويسنده , , M.J. and Vlasov، نويسنده , , Yu. and Iwasaki، نويسنده , , H.، نويسنده ,
Abstract :
In the present work a double ion sensor based on a laser scanned semiconductor transducer (LSST) for the simultaneous determination of K+- and Ca2+-ions in solutions has been developed. Specially elaborated ion-sensitive membrane compositions based on valinomycin and calcium ionophore calcium bis[4-(1,1,3,3-tetramethylbutyl)phenyl] phosphate (t-HDOPP-Ca) were deposited as separate layers on a silanized surface of the Si/SiO2/Si3N4-transducer. The proposed multi-sensor exhibits theoretical sensitivities and the detection limits of the sensor were found to be 2×10−6 mol l−1 for K+ and 5×10−6 mol l−1 for Ca2+. The elaborated double sensor is proposed for the first time as a prototype of a new type of multi-sensor systems for chemical analysis.
Keywords :
Double sensor , Solvent-polymeric membrane , Laser scanned semiconductor transducer (LSST) , Potassium and calcium selectivity