Tanikkul، نويسنده , , Sumalee and Jakmunee، نويسنده , , Jaroon and Rayanakorn، نويسنده , , Mongkon and Grudpan، نويسنده , , Kate and Marquardt، نويسنده , , Brian J and Gross، نويسنده , , Gwen M and Prazen، نويسنده , , Bryan J and Burgess، نويسنده , , Lloyd W and Christian، نويسنده , , Gary D and Synovec، نويسنده , , Robert E، نويسنده ,
Abstract :
A novel Raman sensor using a liquid-core optical waveguide is reported, implementing a Teflon-AF 2400 tube filled with water. An aqueous analyte mixture of benzene, toluene and p-xylene was introduced using a 1000 μl sample loop to the liquid-core waveguide (LCW) sensor and the analytes were preconcentrated on the inside surface of the waveguide tubing. The analytes were then eluted from the waveguide using an acetonitrile–water solvent mixture injected via a 30 μl eluting solvent loop. The preconcentration factor was experimentally determined to be 14-fold, in reasonable agreement with the theoretical preconcentration factor of 33 based upon the sample volume to elution volume ratio. Raman spectra of benzene, toluene and p-xylene were obtained during elution. It was found that analytically useful Raman signals for benzene, toluene and p-xylene were obtained at 992, 1004 and 1206 cm−1, respectively. The relative standard deviation of the method was 3% for three replicate measurements. The limit of detection (LOD) was determined to be 730 ppb (parts per billion by volume) for benzene, exceptional for a system that does not resort to surface enhancement or resonance Raman approaches. The Raman spectra of these test analytes were evaluated for qualitative and quantitative analysis utility.