• Title of article

    Ultra-simple adaptor to convert batch cells with mercury drop electrodes in voltammetric detectors for flow analysis

  • Author/Authors

    Pedrotti، نويسنده , , Jairo J. and Gutz، نويسنده , , Ivano G.R. Gutz، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2003
  • Pages
    11
  • From page
    695
  • To page
    705
  • Abstract
    A simple, robust and fast-responding flow adaptor for mercury drop electrodes (MDEs) is described. An L-shaped PTFE tube with an internal diameter of 0.5 mm is fixed with a silicone ring on the glass capillary of a MDE, in such a way as to direct the outcoming flow onto the mercury drop, from a distance of about 0.5 mm. Any commercial or laboratory-made batch cell, provided with an MDE, serves for the purpose. The level of supporting electrolyte in the cell is maintained constant through a siphon or a lateral draining orifice. The adaptor is compatible with all the different brands and operating modes of the MDEs (free dropping, controlled drop time, renewable static drop, hanging drop or sessile drop). Flow injection experiments were conducted with the following amperometric detection modes: sampled-DC, reverse pulse amperometry (RPA), and anodic stripping voltammetry (ASV). The FIA-RPA peaks presented a R.S.D.<0.8% for 1.0×10−5 mol l−1 lead(II) (N=30, Vsample=100 μl). The response time (0–63% of the signal maximum) to a concentration step is 1.2 s for 500 μl injections of 0.1 mmol l−1 ascorbic acid in acetate buffer at a flow rate of 1 ml min−1, which corresponds to a response volume of 20 μl. As an example of practical application, copper(II) was determined in fertilizers by RPA using the standard addition method, at an analytical frequency of 90 injections per h.
  • Keywords
    Amperometry , Flow cell , Flow injection analysis , Mercury drop electrode , Copper Determination , fertilizers
  • Journal title
    Talanta
  • Serial Year
    2003
  • Journal title
    Talanta
  • Record number

    1644582