Title of article :
A PVC-based 1,8-diaminonaphthalen electrode for selective determination of vanadyl ion
Author/Authors :
Akbari، نويسنده , , A. Shahmohammadi Mousavi، نويسنده , , M.F. and Shamsipur، نويسنده , , M. S. Rahmanifar، نويسنده , , M.S.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2003
Pages :
7
From page :
853
To page :
859
Abstract :
A PVC membrane vanadyl (VO2+) ion-selective electrode was constructed using 1,8-diaminonaphthalen (DAN) as a neutral carrier. The electrode shows good Nernstian response for VO2+ ions over a wide concentration range (1.0×10−1–1.0×10−5 M). The optimum composition of the membrane was 55 wt.% poly(vinylchloride), 35 wt.% 2-nitrophenyl octyl ether (NPOE), 5 wt.% ionophore, and 5 wt.% potassium tetrakis(p-chlorophenyl)borate (KTpClPB). It has relatively fast response time and can be used at least for 5 weeks without any considerable divergence in potentials. The proposed electrode revealed relatively good selectivity for VO2+ over wide variety of other metal ions. The electrode was used for the potentiometric titration of VO2+ ions with EDTA.
Keywords :
Determination , Potentiometeric sensor , Vanadyl ion