Title of article
A study of bismuth-film electrodes for the detection of trace metals by anodic stripping voltammetry and their application to the determination of Pb and Zn in tapwater and human hair
Author/Authors
Kefala، نويسنده , , G and Economou، نويسنده , , A and Voulgaropoulos، نويسنده , , A and Sofoniou، نويسنده , , M، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2003
Pages
8
From page
603
To page
610
Abstract
This work reports the simultaneous determination of Cd(II), Pb(II) and Zn(II) at the low μg l−1 concentration levels by square wave anodic stripping voltammetry (SWASV) on a bismuth-film electrode (BFE) plated in situ. The metal ions and bismuth were simultaneously deposited by reduction at −1.4 V on a rotating glassy carbon disk electrode. Then, the preconcentrated metals were oxidised by scanning the potential of the electrode from −1.4 to 0 V using a square-wave waveform. The stripping current arising from the oxidation of each metal was related to the concentration of each metal in the sample. The parameters for the simultaneous determination of the three metals were investigated with the view to apply this type of voltammetric sensor to real samples containing low concentrations of metals. Using the selected conditions, the limits of detection were 0.2 μg l−1 for Cd and for Pb and 0.7 μg l−1 for Zn at a preconcentration time of 10 min. Finally, BFEʹs were successfully applied to the determination of Pb and Zn in tapwater and human hair and the results were in satisfactory statistical agreement with atomic absorption spectroscopy (AAS).
Keywords
Square wave anodic stripping voltammetry , Bismuth film electrodes
Journal title
Talanta
Serial Year
2003
Journal title
Talanta
Record number
1645039
Link To Document