• Title of article

    A simple chemiluminescence method for determination of chemical oxygen demand values in water

  • Author/Authors

    Hu، نويسنده , , Yonggang and Yang، نويسنده , , Zeyu، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2004
  • Pages
    6
  • From page
    521
  • To page
    526
  • Abstract
    In this paper, a low cost chemiluminescence detector with a photodiode instead of photo-multiplier tube (PMT) was developed for environmental monitoring of water quality. Based on the chemiluminescent reaction of luminol–H2O2–Cr3+ system, light emission caused by luminol–H2O2–Cr3+ system was detected by the photodiode, and its intensity caused by the appearance of Cr3+ after samples digestion was proportional to the chemical oxygen demand (COD). Effects for COD determining such as pH, concentrations, interference, and digestion procedures were investigated. The experimental results show that this suggested method uses an instrument that is simpler and cheaper than the previous ones used for the same purpose. The data obtained by the present method were fairly in good agreement with those obtained by the standard reflux titrimetric method. It has been applied to determine real samples with satisfactory results.
  • Keywords
    Chemiluminescence , Chemical oxygen demand , photodiode
  • Journal title
    Talanta
  • Serial Year
    2004
  • Journal title
    Talanta
  • Record number