• Title of article

    Microwave-assisted extraction and ultrasonic slurry sampling procedures for cobalt determination in geological samples by electrothermal atomic absorption spectroscopy

  • Author/Authors

    Felipe-Sotelo، نويسنده , , M and Carlosena، نويسنده , , A and Fern?ndez، نويسنده , , E and L?pez-Mah??a، نويسنده , , P and Muniategui، نويسنده , , S and Prada، نويسنده , , D، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2004
  • Pages
    8
  • From page
    735
  • To page
    742
  • Abstract
    Slurry sampling is compared to microwave-assisted acid digestion for cobalt determination in soil/sediment samples by electrothermal atomic absorption spectrometry (ETAAS). Furnace temperature programs and appropriate amounts of three chemical modifiers were optimised in order to get the highest signals and good separations between the atomic and background signals. Using nitric acid (0.5% (v/v)) as liquid medium, no chemical modifier was needed. The detection limit, based on integrated absorbance, was 0.04 μg g−1 for digests and slurries. Within-batch precision and analytical recoveries were satisfactory for both procedures. Accuracy was tested by analysing a reference soil and a sediment from IRMM. The methods were further compared employing a set of roadside soils and estuarine sediments. As no significant differences (95% confidence) were found, practical analytical properties were suggested in order to select one of them.
  • Keywords
    Chemical modifiers , Electrothermal AAS , Soils and sediments , Cobalt , Microwave-assisted acid digestion , Ultrasonic slurry sampling
  • Journal title
    Talanta
  • Serial Year
    2004
  • Journal title
    Talanta
  • Record number