Title of article :
Application of Doehlert matrix to the study of flow injection procedure for selenium (IV) determination
Author/Authors :
Hellal، نويسنده , , F. and Dachraoui، نويسنده , , M.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2004
Pages :
6
From page :
1089
To page :
1094
Abstract :
A method for the rapid determination of selenium is reported. It is based on the injection of selenium sample into a stream of KI and subsequent measurement of the increase of the absorption at 290 nm due to the formation of I3−. A Doehlert matrix was applied for the determination of the optimal working conditions. With the developed method, a linear calibration curve from 1 to 10 ppm was obtained. For the analysis of semiconducting thin layers of CuInSe2 (CIS), a cationic resin was added to the sample solution in order to eliminate the effect of copper and indium on the absorbance values. With the developed procedure a sample throughput of 36 samples per hour and a relative standard deviation 1.1% was achieved.
Keywords :
FI , Selenium , Doehlert matrix , CUINSE2