Use of the internal standardization for difficult sampling by graphite furnace atomic absorption spectrometry
Author/Authors :
Oliveira، نويسنده , , Adriana P. and Neto، نويسنده , , José A.Gomes and Nَbrega، نويسنده , , Joaquim A. and Oliveira، نويسنده , , Pedro V.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2004
Pages :
4
From page :
334
To page :
337
Abstract :
This work shows the potentiality of As as internal standard to compensate errors from sampling of sparkling drinking water samples in the determination of selenium by graphite furnace atomic absorption spectrometry. The mixture Pd(NO3)2/Mg(NO3)2 was used as chemical modifier. All samples and reference solutions were automatically spiked with 500 μg l−1 As and 0.2% (v/v) HNO3 by the autosampler, eliminating the need for manual dilutions. For 10 μl dispensed sample into the graphite tube, a good correlation (r=0.9996) was obtained between the ratio of analyte absorbance by the internal standard absorbance and the analyte concentrations. The relative standard deviations (R.S.D.) of measurements varied from 0.05 to 2% and from 1.9 to 5% (n=12) with and without internal standardization, respectively. The limit of detection (LD) based on integrated absorbance was 3.0 μg l−1 Se. Recoveries in the 94–109% range for Se spiked samples were obtained. Internal standardization (IS) improved the repeatability of measurements and increased the lifetime of the graphite tube in ca. 15%.
Keywords :
Sparkling waters , Internal standardization , Graphite furnace atomic absorption spectrometry , Complex matrices
Journal title :
Talanta
Serial Year :
2004
Journal title :
Talanta
Record number :
1646397
Link To Document :
بازگشت