Title of article
Determination of glyphosate mono-isopropylamine salt in process samples using flow injection analysis with tris(2,2′-bipyridyl)ruthenium(II) chemiluminescence detection
Author/Authors
Adcock، نويسنده , , Jacqui L. and Barnett، نويسنده , , Neil W. and Gerardi، نويسنده , , Richard D. and Lenehan، نويسنده , , Claire E. and Lewis، نويسنده , , Simon W.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2004
Pages
4
From page
534
To page
537
Abstract
The mono-isopropylamine salt of glyphosate was selectively determined directly in industrial and commercial formulations using flow injection analysis with tris(2,2′-bipyridyl)ruthenium(II) chemiluminescence detection without the need for separation. Glyphosate and its mono-isopropylamine salt furnished detection limits of 7×10−9 and 3.5×10−10 M and relative standard deviations of 0.4% at 1×10−7 M and 0.8% at 5×10−8 M, respectively. The methodology is robust and reliable with samples subjected only to aqueous dilution prior to analysis.
Journal title
Talanta
Serial Year
2004
Journal title
Talanta
Record number
1646487
Link To Document