Title of article
Integration of a flow-type chemiluminescence detector on a glass electrophoresis chip
Author/Authors
Su، نويسنده , , Rongguo and Lin، نويسنده , , Jin-Ming and Uchiyama، نويسنده , , Katsumi and Yamada، نويسنده , , Masaaki، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2004
Pages
6
From page
1024
To page
1029
Abstract
A glass electrophoresis microchip integrated a flow-type chemiluminescence (CL) detection cell has been developed and evaluated. The chip pattern is a double-T-type electrophoretic sample injection and separation combining with a Y-type chemiluminecent detector. The double-T geometry allows for high-efficiency sample injection and geometric definition of sample plug size. The branch of Y was used as CL reagent channel, and the CL reagent was delivered by a lab-made micropump. Bis[(2,4,6-trichlorophenyl)]oxalate-H2O2 CL system was employed to detect dansyl amino acids. On this microchip, dansyl-phenylalanine and -sarcosine were successfully separated by electrophoresis and detected within 250 s. The detection limits (S/N=3) of dansyl-phenylalanine and -sarcosine could reach to 2.8 and 3.2 μM, respectively, due to the vigorous dilution of sample with CL reagent and timely removal of the waste solution from reaction area.
Keywords
Microchip electrophoresis , Chemiluminescence detection , Flow-type detection cell , Double-T injection
Journal title
Talanta
Serial Year
2004
Journal title
Talanta
Record number
1646737
Link To Document