Title of article :
Sequential injection spectrophotometric determination of trace amounts of iodide by its catalytic effect on the 4,4′-methylenebis(N,N-dimethylaniline)-chloramine-T reaction
Author/Authors :
Tesfaldet، نويسنده , , Zeriet O. and van Staden، نويسنده , , Jacobus F. and Stefan، نويسنده , , Raluca I.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2004
Pages :
7
From page :
1213
To page :
1219
Abstract :
A simple and sensitive sequential injection spectrophotometric procedure is proposed for the determination of trace amounts of iodide in pharmaceutical preparations. The method is based on the catalytic effect of iodide on the (tetra base) 4,4′-methylenebis(N,N-dimethylaniline)-chloramine-T reaction in acidic solution. The method involves a sequential aspiration of 255 μl sample/standard followed by 170 μl tetra base and then 128 μl chloramine-T solutions into a carrier stream to be stacked inside a holding coil and flow reversed through a reaction coil towards a detector. The resulting colored compound is measured at 600 nm using an UV/Vis-spectrophotometer. All the parameters that affect the reaction were evaluated and the calibration curve is linear over a range of 0.1–6.0 μg l−1 of iodide concentration with detection limit of 0.05 μg l−1. A sample throughput of 80 samples per hour and relative standard deviation of less than 2.0% was achieved. The method is successfully applied for the determination of iodide in three different samples (tablets).
Keywords :
Spectrophotometric determination , Sequential injection , reaction
Journal title :
Talanta
Serial Year :
2004
Journal title :
Talanta
Record number :
1646820
Link To Document :
بازگشت