• Title of article

    Flow injection in-valve-mini-column pretreatment combined with ion chromatography for cadmium, lead and zinc determination

  • Author/Authors

    Tanikkul، نويسنده , , Sumalee and Jakmunee، نويسنده , , Jaroon and Lapanantnoppakhun، نويسنده , , Somchai and Rayanakorn، نويسنده , , Mongkon and Sooksamiti، نويسنده , , Ponlayuth and Synovec، نويسنده , , Robert E. and Christian، نويسنده , , Gary D. and Grudpan، نويسنده , , Kate، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2004
  • Pages
    6
  • From page
    1241
  • To page
    1246
  • Abstract
    A flow injection (FI) in-valve-mini-column packed with Chelex-100 resin is proposed for on-line sample pretreatment for some metal ions, namely, Cd(II), Pb(II) and Zn(II), prior to simultaneous determination using ion chromatography (IC). A solution containing a mixture of the cations was first passed through the in-valve-mini-column, followed by on-line elution. The eluate was then flowed further to an injection valve and was injected into an ion chromatograph. Conditions of the system were optimized. A single standard calibration was possible. The recoveries of cations were found to be in the range of 95–105%. The developed method was applied to the accurate analysis of zinc ore samples.
  • Keywords
    Sample pretreatment , In-valve-mini-column , Flow Injection , Ores , Lead , Ion chromatography , Cadmium , Zinc
  • Journal title
    Talanta
  • Serial Year
    2004
  • Journal title
    Talanta
  • Record number

    1646840