Title of article :
Flow injection in-valve-mini-column pretreatment combined with ion chromatography for cadmium, lead and zinc determination
Author/Authors :
Tanikkul، نويسنده , , Sumalee and Jakmunee، نويسنده , , Jaroon and Lapanantnoppakhun، نويسنده , , Somchai and Rayanakorn، نويسنده , , Mongkon and Sooksamiti، نويسنده , , Ponlayuth and Synovec، نويسنده , , Robert E. and Christian، نويسنده , , Gary D. and Grudpan، نويسنده , , Kate، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2004
Pages :
6
From page :
1241
To page :
1246
Abstract :
A flow injection (FI) in-valve-mini-column packed with Chelex-100 resin is proposed for on-line sample pretreatment for some metal ions, namely, Cd(II), Pb(II) and Zn(II), prior to simultaneous determination using ion chromatography (IC). A solution containing a mixture of the cations was first passed through the in-valve-mini-column, followed by on-line elution. The eluate was then flowed further to an injection valve and was injected into an ion chromatograph. Conditions of the system were optimized. A single standard calibration was possible. The recoveries of cations were found to be in the range of 95–105%. The developed method was applied to the accurate analysis of zinc ore samples.
Keywords :
Sample pretreatment , In-valve-mini-column , Flow Injection , Ores , Lead , Ion chromatography , Cadmium , Zinc