Evaluation of on-line preconcentration and flow-injection amperometry for phosphate determination in fresh and marine waters
Author/Authors :
Udnan، نويسنده , , Y. and McKelvie، نويسنده , , I.D. and Grace، نويسنده , , M.R. and Jakmunee، نويسنده , , J. and Grudpan، نويسنده , , K.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2005
Abstract :
Dissolved reactive phosphorus (DRP) was determined as orthophosphate (PO4-P) in fresh and saline water samples by flow-injection (FI) amperometry, without and with in-valve column preconcentration. Detection is based on reduction of the product formed from the reaction of DRP with acidic molybdate at a glassy carbon working electrode (GCE) at 220 mV versus the Ag/AgCl reference electrode. A 0.1 M potassium chloride solution was used as both supporting electrolyte and eluent in the preconcentration system. For the FI configuration without preconcentration, a detection limit of 3.4 μg P l−1 and sample throughput of 70 samples h−1 were achieved. The relative standard deviations for 50 and 500 μg P l−1 orthophosphate standards were 5.2 and 5.9%, respectively. By incorporating an ion exchange preconcentration column, a detection limit of 0.18 μg P l−1 was obtained for a 2-min preconcentration time (R.S.D.s for 0.1 and 1 μg P l−1 standards were 22 and 1.0%, respectively). Potential interference from silicate, sulfide, organic phosphates and sodium chloride were investigated. Both the systems were applied to the analysis of certified reference materials and water samples.
Keywords :
Flow-injection amperometry , Preconcentration , Bio-Rad AG-X8 , Schlieren effect , Phosphate , DRP