Title of article
Hydride generation-headspace single-drop microextraction-electrothermal atomic absorption spectrometry method for determination of selenium in waters after photoassisted prereduction
Author/Authors
Fragueiro، نويسنده , , Sandra and Lavilla، نويسنده , , Isela and Bendicho، نويسنده , , Carlos، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2006
Pages
6
From page
1096
To page
1101
Abstract
A headspace single-drop microextraction (SDME) method has been developed in combination with electrothermal atomic absorption spectrometry for determination of total inorganic Se and Se(IV). SeH2 is generated in a 40 mL volume closed-vial and trapped onto a Pd(II)-aqueous containing drop that is supported at the needle tip of a high-precision chromatographic syringe. Sample pre-treatment by UV irradiation prior to hydride generation allowed converting Se(VI) into Se(IV), hence, facilitating total Se determination. A 2 IV 6-2 fractional factorial design was employed for characterising the effect of relevant variables over SeH2 trapping. The variables showing the most significant effect were Pd(II) concentration in the drop and extraction time. A preconcentration factor of about 25 is achieved. The limit of detection of Se was 0.15 ng/mL using trapping onto a 3-μL drop and the precision expressed as relative standard deviation was about 3%. The limit of detection could be improved further using repeated sampling of the headspace.
Keywords
Selenium hydride , Pd(II)-containing drop , ETAAS , Headspace single-drop microextraction
Journal title
Talanta
Serial Year
2006
Journal title
Talanta
Record number
1649425