• Title of article

    Determination of fluoroquinolones in urine and serum by using high performance liquid chromatography and multiemission scan fluorimetric detection

  • Author/Authors

    Espinosa-Mansilla، نويسنده , , A. and de la Peٌa، نويسنده , , A. Muٌoz and Gَmez، نويسنده , , D. Gonzلlez and Lَpez، نويسنده , , F. Salinas، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2006
  • Pages
    7
  • From page
    1215
  • To page
    1221
  • Abstract
    A high performance liquid chromatography (HPLC) method has been developed for the simultaneous determination of four fluoroquinolones. The studied compounds have been enoxacin (ENO), norfloxacin (NOR), ofloxacin (OFLO) and enrofloxacin (ENRO). An isocratic elution method, using a mixture of tetrahydrofuran (8%) and phosphate buffer (pH 3.00, 30.0 mM, 92%) as mobile phase, has been developed. Fluorimetric detection, exciting at 277 nm, and multiemission scan (407 nm for ENO, 444 nm for both NOR and ENRO and 490 nm for OFLO) has been used. Detection limits of 500, 14.7, 25.2 and 15.0 ng mL−1 for ENO, NOR, OFLO and ENRO, respectively, have been obtained. The proposed method has been satisfactorily applied to analyze NOR, OFLO and ENRO in human urine and serum samples.
  • Keywords
    fluoroquinolones , Enoxacin , Norfloxacin , Ofloxacin , Enrofloxacin , HPLC , serum , urine
  • Journal title
    Talanta
  • Serial Year
    2006
  • Journal title
    Talanta
  • Record number