Title of article :
Wide range pH measurements using a single H+-selective chromoionophore and a time-based flow method
Author/Authors :
Safavi، نويسنده , , Afsaneh and Rostamzadeh، نويسنده , , Abolfazl and Maesum، نويسنده , , Saeed، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2006
Abstract :
A hydrophilic transparent triacetyl cellulose membrane was adopted as a pH optode by immobilizing highly selective and sensitive Nile blue indicator on the membrane. Contrary to the common procedure for determinations using optodes, in which a steady state response is measured, a new approach is introduced in which the dynamic response of the optode is used as the analytical signal. While in common procedures, pH optodes exhibit limited linear dynamic range (often 2–4 pH units only), it is shown that in a time-based flow method, an optode with only one acid-base indicator can be used for measurement in the pH range of 0–10. The procedure is simple, inexpensive and rapid.
Keywords :
Optode , Wide range pH measurement , Dynamic method