Title of article
An economic approach for construction of a multiplexed capillary electrophoresis system
Author/Authors
Liu، نويسنده , , Shaorong and Pu، نويسنده , , Qiaosheng and Gao، نويسنده , , Lin and Lu، نويسنده , , Joann، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2006
Pages
7
From page
644
To page
650
Abstract
We describe in this report an economic approach to construct a multiplexed capillary electrophoresis (CE) system with optic-fiber UV detection. To demonstrate the feasibility of this approach, a seven-lane CE setup is built. The system has a background noise of ∼10−5 absorbance unit, a limit of detection of 3–4 μM for uracil at 254 nm, and a linear dynamic range of 2.5 orders of magnitude. The apparatus consists of three major components: a UV light source, a photodiode array (PDA) and optical module assembly, and a computer. Compared to the recently reported 96-lane capillary electrophoresis system with UV–vis absorbance detection, the cost of goods is reduced by more than 10-fold, and the noise level is improved by 2 to 3 folds. Parallel CE separations of a protein mixture have been performed on this apparatus, and reproducible and high resolution separation results have been obtained. Detailed construction and characterization of the system have also been discussed.
Keywords
High throughput protein analysis , Capillary array electrophoresis , Optic fiber UV detection , Protein Separation , Multiplexed capillary electrophoresis
Journal title
Talanta
Serial Year
2006
Journal title
Talanta
Record number
1650736
Link To Document