Title of article
Potentiometric measurements in sequential injection analysis lab-on-valve (SIA-LOV) flow-system
Author/Authors
Kikas، نويسنده , , Timo and Ivaska، نويسنده , , Ari، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2007
Pages
5
From page
160
To page
164
Abstract
Advantages of using sequential injection analysis lab-on-valve (SIA-LOV) in potentiometric measurements are studied with Ca2+ sensitive solid-contact ion-selective electrodes (SC-ISE) and pH electrode based on polyaniline (PANI). Experiments show that Ca2+-SC-ISE requires a stopped flow mode to be used in order to get longer equilibration time. On the other hand, PANI based pH electrode gives better results under flow conditions. SIA-LOV was found to be a flexible solution handling system for potentiometric measurements. The technique gives a possibility to fine-tune the calibration range or even recondition the electrode before every measurement without losing in linear range of calibration.
Keywords
Sequential injection , Lab-on-valve , Ion-selective electrodes